CLINICAL GUIDANCE | SECTION ONE
7
1.1.1 Behavioural Procedures
1. Measure impedances by performing an implant test.
2. In the Open or Create MAP screen, create a new MAP by selecting the
required MAP parameters from the available drop-down menus and
clicking Create. Alternatively, you can select and open an existing MAP
that you can modify.
The default streamlined programming method is the behavioural one and
by default, channels 22, 16, 11, 6, and 1 are highlighted for measurement.
Any flagged channels are noted.
3. Click on the Streamlined Programming drop-down menu. Using the
“Set Ts” buttons, measure the T-level on each of the default channels
(indicated by wider bars) using an ascending method. By default, the
starting level is 80 CL. For more information, see Programming
Principles T-level Assessment Procedure. Custom Sound
automatically interpolates T-levels on the unmeasured channels
(indicated by the thinner bars).
4. At initial activation, use of streamlined programming methods is
recommended. However, if a large variability in the profile is observed,
additional measurements should be made in this area of the array, for
example on the basal electrodes (e.g. electrodes 4, 3 and 2) to improve
the accuracy of the interpolation and obtain optimum sound quality.
5. To set C-levels, click the “go live” button (the large green button) and
talk to the recipient as you gradually increase C-levels globally by
clicking the <up> arrow under “live Cs”. During live voice testing, the
microphone sensitivity is set to 12 by default for most processors (the
optimal sensitivity) and the volume control is set to maximum (this can
be reduced). Adjust C-levels so that ongoing conversational speech is
comfortably loud. It is recommended reading aloud to the recipient and
allowing the person to follow along with the text while tracking loudness
using the loudness-scaling chart. As the loudness level gets closer to
comfortably loud, reduce the step size from 5 CL to 2-3 CL. If necessary,
you can reduce the microphone sensitivity slightly.
6. Note: When going live to set C-levels, ADRO
®
is typically used, however it
can be turned off when setting C-levels using live voice if desired.
If you choose to use the traditional method of measuring individual C-levels
and then testing in live speech mode, it is recommended that you initially
reduce the microphone sensitivity or volume if it is at maximum because the
sound could be too loud. Gradually increase the sensitivity up to the default of
12 and check sound comfort at maximum volume.