CLINICAL GUIDANCE | SECTION ONE
35
softer speech sounds. In both cases, C-levels should be re-assessed.
The Sensitivity Mode may be set to ‘ManualorFixed’. When set to the ‘Manual
mode the recipient can manually adjust the sensitivity setting by pressing the
appropriate buttons on the Freedom
. This can also be done with the CP810
Sound Processor if the Advanced (sensitivity)UI (user interface) is selected in
Custom Sound
or the CR110 Remote Assistant can be used to make sensitivity
changes on the CP810 processor. When set to the ‘Fixedmode (or the ‘Simple
UI is indicated for the CP810 processor), the sound processor buttons to adjust
sensitivity do not function. (The sensitivity can also be locked in the CR110
Remote Assistant used with the CP810 processor by choosing this option in
Custom Sound.) The fixed mode is sometimes the preferred mode for the
programming of young children, where it is preferable that the sensitivity is not
adjusted. In this case it is recommended to also choose an environment setting
that includes autosensitivity
(ASC). The function of the autosensitivity control
(ASC) is to automatically adjust the microphone sensitivity based on the noise
oor of the incoming acoustic signal. If the noise floor is above a programmable
breakpoint, it is likely that the gain (sensitivity) is too high for that situation. In
this case the ASC gradually reduces the sensitivity. If the noise floor is below
the breakpoint the gain is gradually increased back to the manual setting. The
programmable breakpoint is referred to as theAutosensitivity Breakpoint’ and
is by default set to 57 dB SPL for the CP810/Freedom processors. This means
that the autosensitivity function is activated (i.e. will start turning down the
sensitivity), once the background noise reaches 57 dB SPL.
ASC is available as a setting for all Cochlear sound processors.
1.11.2 Volume Setting
The CP810 and Freedom
processors provide the option to adjust the volume
level in addition to the microphone sensitivity. For the CP810 processor,
the maximum volume setting display is 10 and the minimum setting is 1.
For Freedom, the maximum volume setting display for 9, and the minimum
setting is 0. The effect of adjusting the volume setting is to globally reduce
the C-levels by a certain percentage of the dynamic range. The overall range
of reduction is programmable. The default setting involves a 20% reduction in
C-levels between the maximum and minimum volume settings.The maximum
volume level results in stimulation at C-level. If the recipient requires the option
to increase the volume under certain listening situations, it is recommended
that a global increase in C-levels be applied to the MAP, and that the volume
setting be reduced to an intermediate setting to provide room to adjust the
volume in both directions.