SECTION TWO | CLINICAL GUIDANCE
60
Scenario One
In the best case scenario, all channels would have both T-levels and C-levels
below any side-effect threshold (Figure 4).
Scenario Two
In the worst case, it is impossible to find T-levels and C-levels since they are
above the threshold of a side effect on all channels (Figure 5). This implies
that the device may not be positioned optimally over the cochlear nucleus, or
the cochlear nucleus is not responsive to electrical stimulation (e.g. irradiation
damage).
Figure 4 : Best case
ABI programming
scenario
Figure 5 : Worst case
ABI programming
scenario