SECTION TWO | CLINICAL GUIDANCE
64
Table 3: Maximum safe current and corresponding current levels for various
stimulus pulse widths for the Nucleus ABIM implant
Pulse width
(µs)
Maximum safe
Current Level (CL)
25 >255
50 >255
100 >255
150 >255
200 255
250 244
300 235
350 227
400 221
The above table shows that up to pulse widths of 200 µs, the full stimulator
output (255 units) may be safely used. Beyond this, the inverse proportional
relationship between pulse width and maximum safe current imposes Current
Level values progressively less than 255 with increasing Pulse Widths. To ensure
that you remain within established safety limits, please refer to Table 2 above.
Scenario Four
In rare circumstances it may be possible to obtain T-levels on several channels,
yet not achieve C-level on any channel before a side effect occurs, as shown in
Figure 8.
In this case, see Managing side effects. If this does not improve the situation, it
will probably be ineffective to balance these channels to an overall significantly
reduced C-level. This may potentially limit the detection of temporal
information in speech.
Figure 8 : T-levels
found with C-levels
not achieved before
a side effect occurs