
SECTION ONE | CLINICAL GUIDANCE
50
• Alternating stimulus polarity will effectively reduce the stimulus artefact.
However, there may be a disadvantage if the evoked brainstem responses
from negative and positive leading stimulus pulses have different
response latencies. This can be reduced by using narrow pulse widths, for
example, 25 μs.
(Absence of an EABR does not mean that the implant is not working or that
the individual will not stimulate postoperatively.)
1.14.2 Cortical Evoked Potentials (CEP)
Cortical Evoked potentials can be recorded in a similar manner as EABR.
Practically, the only difference between EABR and CEP is the time between the
stimulation pulse and the recording window. Typically, the EABR signal occurs
at 1 – 10 ms after the stimulus while the CEP signals can be recorded between
100 and 500 ms. Clinically, CEP is usually more associated with perception
(conscious), while EABR is merely a sign of detection (unconscious).
1.14.3 Electrical Stapedius Reflex Test (ESRT)
The acoustical stapedius reflex is a small movement of the stapedius muscle in
the middle ear that is elicited by loud (>80 dB SL) acoustical stimulation. The
ESRT can be tested with an implant recipient and is generally believed to have
a correlation with C-level. There are two ways to measure ESRT thresholds:
Method 1: Visual inspection ESRT
During surgery, after the electrode has been placed and with the skin flap still
open, the implant can be stimulated and the surgeon can look for a stapedius
reflex by visual inspection of the stapedius tendon or the middle ear ossicles.
Please keep in mind the following tips:
• Make sure the monopolar reference electrodes are covered to get a good
electrical connection. Preferably test the impedance first. If needed, use
saline to improve the electrical contact.
• Do not use a muscle-relaxing agent since this will suppress the ESRT.
Method 2: Impedance bridge testing of ESRT
When using this method, an electro acoustic immittance meter is connected
to either the contra- or the ipsilateral ear. By measuring the impedance change
during the electrical stimulation, a response can be observed. To use this
method, use your immittance meter in free running mode and observe for
changes in the impedance when stimulating. A trigger is also made available
by the system, should you use equipment that needs to be triggered for
measurements. This trigger functionality is similar to the description given for
EABR measurements.
Comentários a estes Manuais