SECTION ONE | CLINICAL GUIDANCE
42
The effects of Beam were evaluated in a double-blind trial with five adult
Nucleus
®
CI users
4
.
Subjects were tested with their standard program and
were then fitted with Beam. Testing with Beam was carried out after a two-
week trial period. Speech Reception Threshold (SRT) with sentences and the
percentage correct phoneme scores for Consonant-Vowel-Consonant (CVC)
words were measured in quiet and noise at different signal-to-noise ratios.
The results are shown in quiet (quiet), speech weighted noise at 90° (spw 90),
multi-talker babble at 90° (babble 90), speech weighted noise from three
directions 90°, 180° and 270° (spw 3n), and multi-talker babble from three
directions 90°, 180° and 270° (babble 3n). Beam showed significant benefits
for listening to speech in competing noise. As shown in Figure 15, the group
showed an average improvement of 5 – 16 dB in SRT in noise when wearing
Beam compared to their standard program. Percentage correct phoneme
scores improved from 10 – 41 percentage points.
Focus with the CP810 Sound Processor
The CP810 utilizes two programmable omni-directional microphones, with
long filters and individual customized calibration of up to 1 dB precision, as well
as proprietary DSP technology that combines them into selectable response
patterns. In addition to providing greater noise cancellation from the rear and
sides than the Freedom processor, CP810 Focus more reliably preserves sound
from in front.
A comparison of Focus with the Freedom sound processor and the CP810
sound processor was conducted as part of the Nucleus 5 validation
5
. Speech-
in-noise testing was conducted with 19 Australian subjects using a spatially
separated noise setup, in which four talker babble was delivered from speakers
positioned at left, right, and behind the subject (N 90, 180, 270). The signal-to-
noise ratio was optimized for each participant to avoid floor and ceiling effects.
Figure 15:
Improvement of
Speech Reception
Threshold (SRT) with
Beam in different
noise scenarios
Vista de página 43