SECTION TWO | CLINICAL GUIDANCE
58
7. Position the transmitter coil so it is just below the retainer disk. The
transmitter coil will then be located directly over the receiver coil.
8. Re-apply retainer disks every one to three days as required.
Finding the receiver/stimulator
To find the receiver/stimulator:
1. Palpate the area or check for a tattoo, (if used).
2. Use Custom Sound software to connect to the receiver/stimulator.
3. Start testing with headband, then shave the scalp for the retainer disk.
Note: To place retainer disk, clean (i.e. with alcohol) and dry a quarter-sized
area and place disk slightly higher than you think is required. Do not perform
an impedance test until all electrode T- and C- levels have been checked.
Preparing the recipient for responses
When programming an ABI recipient, you may want to ask the recipient to
provide the following responses:
Let you know as soon as the recipient hears or feels anything;
Let you know when sound is comfortably loud;
Let you know what it sounds like (i.e. bell, tone, buzz, etc.);
How much the recipient notices the side effects;
Give you a number for the pitch of the sound.
Rating Scales
Pitch magnitude 1 – 100
Non-auditory sensations 1 – 4
Loudness 1 – 10
Figure 3:
Positioning the
retainer disk