SECTION ONE | CLINICAL GUIDANCE
46
Recommendations for changing Autosensitivity (ASC) breakpoint
Increasing the Autosensitivity breakpoint will make it less aggressive / prevent
it from turning down the sensitivity so much. Decreasing the Autosensitivity
breakpoint will make it turn down the sensitivity more. It is not recommended
to change both the Autosensitivity breakpoint and the base level (for the
ESPrit 3G processor). If you change base level breakpoint or Autosensitivity
breakpoint parameters from the default, it will result in a complex interaction
between the two. If you change one, it is recommended to set the other to
the default.
Recommendations for changing Autosensitivity (ASC) breakpoint
Indication
Recommendation
for a CP810 and
Freedom
™
sound
processor
Recommendation
for an ESPrit
™
3G
sound processor
Outcome
Background noises
are too soft or all
sound is becoming
too soft
Increase the
Autosensitivity
breakpoint, for
example from
57 to 59
Increase the
Autosensitivity
breakpoint, for
example from
-15 to -13
Autosensitivity will
be less aggressive
/ prevent from
turning down the
sensitivity too much
Background
noises are too
loud / becoming
troublesome
Decrease the
Autosensitivity
breakpoint, for
example from
57 to 55
Decrease the
Autosensitivity
breakpoint, for
example from
-15 to -17
Autosensitivity
will turn down the
sensitivity more
Autosensitivity is used alone or in combination with other input processing
technologies for the SmartSound Everyday, Noise, Focus and Music programs.
See section 1.13 SmartSound for more information.
1.13 SMARTSOUND
™
ENVIRONMENTS
A SmartSound
™
Environment provides guidance as to which SmartSound
option is best suited to a particular listening environment. Each environment
has a default option and sub-options for differing recipient needs. For a
Freedom
™
MAP, the following SmartSound environments are available. The
default option for an environment is highlighted in bold font.
Comentários a estes Manuais